• DryShield Sterilization

  • The DryShield DS1 is made from a high heat-tolerant material, and the DryShield mouthpiece is made of an autoclavable silicone material. All DryShield components are completely autoclavable. DryShield requires the following standard infection control protocols for sterilization:

  • DryShield DS1

    • After completion of dental procedure, run the DryShield DS1 with a non-corrosive, pH-neutral enzymatic cleaning solution and perform a surface disinfection with a disinfectant wipe between each patient. The DryShield DS1 can also be autoclaved.
    • Maintenance of the lever O-ring is done through visual inspection. The O-ring should be replaced when there's a noticeable decrease in lever function.

  • DryShield Mouthpiece

    • After completion of each dental procedure, detach mouthpiece from bite block and place both in ultrasonic cleaner for 15 minutes.
    • Next, clean the separated mouthpiece and bite block of debris by thoroughly scrubbing with soap and water or a detergent solution.
    • Clean the inside of the mouthpiece through access at the slit opening and at the suction connector region using a long-handled brush and thorough scrubbing.
    • Then, rinse under running water, air dry and place in autoclave pouch to be sterilized.

    • The DryShield mouthpiece MUST BE sterilized between each use. Sterilize in a steam autoclave machine at minimum 121 °C (250 °F) for a minimum of 30 minutes. Allow a minimum of 30 minutes in drying cycle.

  • Important Notes

    • DO NOT cold-sterilize DryShield DS1 or mouthpieces.
    • Always follow DryShield sterilization protocol.